Products Introduction:
The microcomputer analog controls the infrared thermal radiation, drug ion delivery and electric pulse to work on the lesion tissue, speeding up the metabolism, activating immunity system, regulating excitement of nerve ending, intensifying drug concentration to achieve maximal therapy effect. It can be used in the treatment of mastitis, hyperplasia of mammary glands, mastodynia, benign lump, irregular menstruation and imbalance of hormone level. Special curing effect is found in breast health care and stimulating milk secretion.
1. The electricity pulse therapy has seven kind of different waves: We can choose different therapy method according to different patient or different spot.
2. Continuous adjustable Infrared intensity range: band 1-18, intensity increasing by degree gradually.
3. The medicine ion inducts: We can directly put the medicine into two electrodes surfaces, cause the medicine directly close to human body, it can play the better assistance treatment role.
This instrument has time alarm function, error alarm function, infrared radiation intensity alarm function and so on.
This instrument has the remarkable curative effect on mastitis, mammary gland proliferation and so on.
Specification:
1. Seven optional therapy modes: condensation wave, spike wave, rarefaction wave, fluctuation wave, intermittent wave, spiccato fluctuation wave and variable frequency wave.
2. Input power: ≤ 20VA
3. Continuous adjustable Infrared intensity range: band 1-18, intensity increasing by degree gradually.
4. Probe wave length: 0.8μm-3μm.
5. Continuous adjustable Timing range: 0-99min, time error ≤±1min.
The performance of electricity stimulation:
1. Seven optional therapy modes: condensation wave, spike wave, rarefaction wave, fluctuation wave, intermittent wave, spiccato fluctuation wave and variable frequency wave. Corresponding voltage output peak-refer to the peak value of diagram 1.
2. When the output scope of instrument is the biggest, each pulse electric quantity should be bigger than 7μC.
3. The effective value should not be bigger than the value on the diagram 2 when the instrument was maximum outputting scope.
4. The single pulse maximum output energy of instrument skin electrode does not surpass 300mJ.
5. When the instrument surveyed under the interruption, the output voltage peak value not to be bigger than 500V.
6. The instrument should withstand the influence which direct current interruption and short-circuits, its performance does not have to weaken.
7. The adjustment of output scope should be continuously even.
8. When the output frequency of instrument is bigger than 1000Hz, the change of output current rate should not be bigger than 5%.
Regular working conditions:
1. Ambient temperature: 5℃~40℃;
2. Relative humidity:≤85% ;
3. Atmospheric pressure:86kPa~106kPa;
4. Power source:AC220V±22V 50Hz±1Hz
5. Gross Weight: 3kg; Dimension: 450×350×250mm