Performance |
Display: 12"color TFT display to show 4~9 waveforms |
Resolution: 640×480 or 800×600 |
Interface: interface for CRT display and net card |
Tend observation: Tend chart of 1 – 72 hours |
Alarm: upper and lower limit alarm, audible and visual, third grade alarm |
Connection: Can be connected to central monitor and long-distance monitor |
Memory method: manual, alarm, timing, etc. |
Optional configuration: printer |
|
ECG |
Configuration: 5 lead |
Lead input: 5 lead:RA,LA,RL,LL,V |
Lead Selection: Ⅰ,Ⅱ,Ⅲ,avR,avL,avF,V,MCL1 |
Gain: 4 grades optional |
Sweep speed: 3 grades optional |
Leakage of current: less than 3μA |
Safety: endurance of voltage of 4000VAC/50Hz,defibrillation protection and contra electrotome interface |
Indication for electrode cast: yes |
Work mode: diagnosis, monitor,and surgery |
Alarm indication: audible and visual, and can be reviewed |
ST segment monitor: yes, and alarm range can be set. |
|
Heart rate |
Measuring range: 25 to 250 bpm |
Accuracy: ±2 bmp |
Tendency chart: 1~24 hours |
Refurbishing rate: once for each beat |
Alarm range: sustained adjustment of upper and lower limit between 20~250 |
|
Respiration |
Measuring range: 1~99 bmp |
Measuring error: ±2bmp |
Measuring method: RA-LL Impedance |
Display: waveform, respiratory rate |
Alarm: yes |
Alarm range: adjustable |
|
SpO2 |
Measuring range: 0~100% |
Accuracy: 80~100%:±2%,60~79%:±3%,and have no requirements when less than 60% |
Sensor type: double wavelength LBD |
Resolution: ±1% |
Alarm range: sustained adjustment of upper and lower limit between 1~100% |
Alarm indication: audible and visual |
Pulse rate monitor: yes |
Pulse rate range: 0~255bmp |
|
Non-invasive pulse pressure |
Pulse rate range: automatic vibration |
Work mode: automatic/manual |
Measuring parameters: SYST/DIAS/MEAN/PR |
Measuring mode: adult, pediatric, infant |
Unit: mmHg or kPa |
Excess pressure safety: double safe guard |
Alarm range: 25~240 mmHg adjustable |
|
Temperature |
Channel: two |
Measuring range: 25~45℃ |
Unit: ℃ or ℉ |
Accuracy: ±0.2℃ |
Alarm range: 25~45℃ |
Alarm indication: yes |