Specifications: |
Test Items |
WBC, LY#, MO#, GR#, LY%, MO%, GR%, HGB, RBC, HCT, MCV, MCH, MCHC, RDW—CV, RDW—SD, PLT, PCT, MPV, PDW, WBC Histogram, RBC Histogram, PLT Histogram |
Background Count |
WBC: ≤0.3×109/L RBC: ≤0.03×1012/L HGB: ≤1g/L PLT: ≤10×109/L |
Linearity |
Test Item |
Test Range |
Allow Difference of Linearity |
WBC |
0∼6.0×109/L |
±0.3×109/L |
6.0∼99.9×109/L |
±5% |
RBC |
0∼0.99×1012/L |
±0.05×1012/ L |
1.0∼9.99×1012/L |
±5% |
HGB |
0∼99 g/L |
±2g/L |
100∼300 g/L |
±2% |
PLT |
0∼99×109/L |
±10×109/L |
100∼999×109/L |
±10% |
Repeatability |
Test Item |
CV(%) |
Teat Range |
WBC |
≤2.0 |
7.0~15.0×109/L |
RBC |
≤0.5 |
3.5~6.0×1012/L |
HGB |
≤1.5 |
110~180g/L |
PLT |
≤4.0 |
150~500×109/L |
MCV |
≤0.5 |
80~110fL |
Sample Volume |
Twig blood: 20μl |
Dilution Ratio |
WBC/HGB: 1:500 RBC/PLT: 1:50000 |
Interfaces |
RS232 joint, USB, PS2, Parallel port |
Test Speed |
60 samples/hours |
Continuous Work Time |
>8h |
|
|
|
|